CHAPITRE 5. ADVANCED FEATURES
F
IGURE
5.4 : Affine conversion settings
In the case of our sensor, when it measures 0 Bar, it delivers 4mA, which corresponds to 0.2V across shunt
(1)
. When it measures 10 Bar, the sensor delivers 20 mA, which corresponds to 1V across shunt
(2).
5.2
|
Alarms setting (available in a later version)
5.2.1
4 logic output (Alarms)
Alarms A, B, C and D send a TTL voltage signal when the set event occurs.
F
IGURE
5.5 : Sorties alarmes
When activating a logic output, you must first define the alarm type
(1)
:
• Trigger condition : activation on event without starting/stopping recording
• Recording start : activation of measurement file recording during the event
5.3
|
Remote control
ò
To use the remote control functions, the device must be connected to a network. Please refer to
the "Network settings" chapter for further information on configuration.
DAS
User manual
46