66
Control solution
NOTE
Use only Omnitest® 5 Control for
the Omnitest® 5 monitoring system.
The control solution is sold
separately.
Verification with control solution
The control solution Omnitest® 5 Control verifies that the blood
glucose monitoring system is functioning properly.
The control solution should be used in the following cases:
For any suspected malfunction of the device or test strips
If your blood glucose readings do not match your symptoms or
if you suspect that the readings are not accurate
If the meter was dropped
For quality control in professional applications
For training purposes and when learning to operate the
system