32
Applying blood
Touch your finger to the tip of the Omnitest® 5 test strip.
Ensure that blood is not applied from above.
The measurement chamber of the test strip will draw automa-
tically the blood of your finger. Your finger should remain still,
until the confirmation window is completely filled. The minimal
sample volume is 0.5 μL.
NOTE
Do not squeeze the fingertip. Do not attempt to apply a
smeared blood sample. The measuring chamber must be filled
in one draw. Reapplying can lead to incorrect readings.
Do not force your finger against the test strip, the test strip
must not be bent.
Measuring blood glucose
Summary of Contents for Omnitest 5