A
ppe
n
d
ix
81
Problem
Cause
What to do
Different readings in
comparison to another
device.
The other device has been calibrated
to whole blood while the Omnitest® 5 is
calibrated to plasma.
To check the accuracy of your system, please use the control
solution Omnitest® 5 Control or compare readings to a laboratory
device that works with plasma.
The blood samples were not collected
at the same time.
Repeat the test and take the blood samples at the same time
for comparison.
The puncture site was not cleaned.
Repeat the test with a blood sample from a clean puncture site.
The blood sample was smeared or could
not be taken in one draw from the test
strip.
Repeat the test with a sufficiently large fresh blood sample.
The meter does not
turn on.
The battery voltage is not adequate or
there is a problem with the meter.
Change the batteries. If the problem persists, contact your retailer
or B. Braun representative in your area.
After applying the sample
the test does not start.
The sample volume is not adequate or
there is a problem with the test strips or
the meter.
Apply a sufficient sample volume to the new test strip and repeat
the test. Perform a test with the control solution or the check strip.
The result is doubtful.
There is a potential problem
with the test strips.
Repeat the test with a new test strip. Test with control solution.
The meter code and the code printed
on the test strip vial differ.
Repeat the test with a new test strip. If the problem persists,
contact your retailer or B. Braun representative in your area.
Summary of Contents for Omnitest 5
Page 1: ...Omnitest 5 Blood Glucose Monitoring System ...
Page 2: ...2 ...
Page 6: ...Nutzung 6 Overview ...
Page 18: ...18 Measuring blood glucose ...
Page 44: ...44 Stored readings ...
Page 45: ...45 Contents Displaying stored readings Transferring readings ...
Page 50: ...50 Settings ...
Page 51: ...51 Contents Time and date Acoustic signal Timer Target level Alarms Exit settings ...
Page 61: ...S e t t i n g s 61 ...
Page 62: ...62 Functional check ...
Page 63: ...63 Contents Check strip Control solution ...
Page 72: ...Nutzung 72 Appendix ...