Wipe disinfection for electrical devices
RT: Room
temperature
*
The manufacturer recommends Super Sani-Cloth® Germicidal Disposable Wipe (PDI, EPA register number 9480-4)
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
Phase II
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe the specified application time (2 min minimum).
7.6
Inspection, maintenance and checks
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
►
Set aside the product if it is damaged.
►
Periodically download or delete video files on the system computer to
avoid filling up the storage drive.
The DSM system will not allow the user to make new recordings if the
storage drive is filled up.
7.7
Packaging
►
Ensure that the packaging will prevent a recontamination of the prod-
uct.
7.8
Storage and transport
Fig. 28
►
Ensure that the robotic arm is in “Storage” position before moving the
cart, see Chapter 6.2.1.
►
Wrap the power cable around its wrapping post and hang the wireless
footswitch from its on the front of the cart.Release the brakes on each
of the four casters and move the 3D display into the position illustrated
above.
►
Twist the arm joint handles to lock the arm in place.
►
Lock the cart castors after moving the system.
8.
Maintenance
Preventive maintenance
The Digital Surgical Microscope does not consist of any device that
requires a regularly maintenance within a defined time frame.
However, in an interval of 12 month a regular repeat inspection is manda-
tory. The repeat inspection has also to be carried out after each kind of
repair, if the device or one component was dropped or damaged or if the
medical device was not used according to its intended use.
CAUTION
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
►
Use cleaning and disinfecting agents according
to the manufacturer’s instructions which
–
are approved for plastic material and high-
grade steel,
–
do not attack softeners (e.g. in silicone).
►
Observe specifications regarding concentration,
temperature and exposure time.
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water quality
Chemical
I
Cleaning
RT
-
-
-
55% isopropyl alcohol,
0,25% n-alkyl dimethyl ethyl benzyl ammonium chloride,
0,25% n-alkyl dimethyl benzyl ammonium chloride*
II
Wipe disinfection
RT
≥
2
-
-
55% isopropyl alcohol,
0,25% n-alkyl dimethyl ethyl benzyl ammonium chloride,
0,25% n-alkyl dimethyl benzyl ammonium chloride*