Guidance and manufacturer's declaration – electromagnetic immunity
The model GT SERIES is intended for use in the electromagnetic environment specified below.
The customer or the user of the model GT SERIES should assure that it is used in such an environment.
Immunity test
IEC 60601test
level
Compliance
level
Electromagnetic environment
–
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 A/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the model
GT SERIES, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d =
1,2
P
d =
1,2 80 MHZ to 800 MHZ
P
d =
2,3 800 MHZ to 2,5 GHZ
P
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
a
electromagnetic site survey, should be
less than the compliance level in each
b
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the model GT SERIES is used exceeds the applicable RF compliance level above, the model GT SERIES
should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the model GT SERIES.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than V/m.
62
ISO 60601-1: Toegepaste deelclassificatie - voldoet aan normen voor
bescherming tegen elektrische schokken voor type B apparatuur
Koelsysteem
:
>>
Als de temperatuur 45°C overschrijdt, zal de ventilator blijven draaien tot de
interne temperatuur is gedaald.
>>
Als de temperatuur hoger is dan 65°C, zal het apparaat automatisch worden
uitgeschakeld en blijft de ventilator draaien. Eens de interne temperatuur is
gedaald, kunt u het apparaat opnieuw opstarten.
>>
Als het apparaat wordt gebruikt voor continue uitharden, verminder dan de
uithardingstijd.
>>
In de veronderstelling dat orale weefsels niet worden aangeraakt wanneer een
uithardingslamp wordt gebruikt of wanneer het behandelde gebied of de
bedieningsinstellingen worden omgeschakeld is het nodig de lamp een 30 seconden
te laten afkoelen voor u de volgende uithardingscyclus begint.
ISO 60417-5009: STAND-BY
IEC 60417-5031: Gelijkstroom
ISO 7000-1641 Volg de gebruiksaanwijzing of raadpleeg instructies voor
gebruik
Verklaring van grafische symbolen
:
Fabricagedatum
Opslagtemperatuur: -20°C ~ 65°C
Fabrikant
Klasse II apparatuur
>>
Dit apparaat is ontworpen met een ingebouwde koelventilator, die automatisch
wordt geactiveerd wanneer het apparaat in gebruik is, en die een paar seconden
na het einde van de cyclus blijft lopen.