24
USER MANUAL
__________________________________________________________________________________________________________________
15.5. Percent calculation functions (PErCEnt)
The function allows displaying
weighing result in percent.
Measurement is performed in two
phases:
-first phase – weighing reference
mass (mass referenced to 100%),
-second phase – weighing any
sample mass as a percent of
reference mass measured in first
phase.
Weighing result is displayed in
various formats, depending on
reference mass value. For reference
mass values between 0÷3,5% of
scale capacity, format of weighing
result is
100.0
, for values between
3,5÷35% it is
100.00
and above
35% -
100.000
.
The function has the following
options:
-
PEr oFF
– switch the function off,
-
PEr on
– set current indication as
100%, show indications in %,
-
out –
exit without changes.
Caution:
1.
PEr Err
message informs that
reference mass is less than 0,5
⋅
Min
or was not defined.
2. When the scale shows weighing
result in percent,
→
T
←
key works
as usual.
0.00 g
19.07 g
PErCEnt
PEr oFF
PEr on
100.00%
50.00%
100%
50%
MENU
T
T