15.7. Function for calibration with external weight (CALIbr)
Calibration with external weight should be performed if scale accuracy is not
satisfactory. Calibration weight stated in technical data table for the scale (or of
better accuracy) should be used then.
Operations sequence:
Press
MENU
key to display user
functions, shown one by one in
loop.
Press
→
T
←
key when
CALIbr
function appears.
The following options will be
displayed:
-
CAL on
– calibration with external
recommanded value weight (see
technical data).
-
CAL StP
– calibration with
external weight, confirmation of
next steps -
MENU
key,
-
CAL Pt
– calibration with external
weight other value,
out
– leave without changes
Press
→
T
←
key when
CAL on
option appears.
Wait for writing zero to the scale.
When
LOAD
message appears put
calibration weight on the pan.
Wait until internal calibration is
finished
zero
indication
is
displayed.
Attention:
CAL StP
option generates commu-nications
Press
Menu.
- CAL -
0.00g
m
LOAD
m
MENU
T
T
0.00g
0.00g
CALIbr
CAL on
CAL oFF
...