USER MANUAL
15
____________________________________________________________________________________________________________________
is installed and on) and nett weight. During transmission
LabEL
communicate is
displayed.
The way of sending data and transmission parameters is set using
SErIAL
special function.
Set of send data is set using special function
.
The following data can be send:
-
Header (scale type, Max, d, e, serial number),
-
Operator identification number,
-
Successive printout number (measurement),
-
Identification number or product bar code,
-
Number of pcs (PCS function only),
-
Single detail mass (PCS function only),
-
Nett weight,
-
Tare (package mass),
-
Gross weight,
-
Total mass (Total function only).
Computer must have a special program for cooperation with data from a scale.
Dedicated programs are also offered by AXIS.
Except RS232C joint, the scale can be equipped with USB or Wi-Fi interface.
Needed controllers and instruction can be found on a CD supplied with Axis
scales.
Detailed protocol description
LonG protocol
Communication parameters: 8 bits, 1 stop bit, no parity, baud rate 4800bps,
initialising signal (data send order) – correspond to press
key
Computer
Scale: S I CR LF (53h 49h 0Dh 0Ah),
Scale
Computer: scale response according to description below (16 bytes):
Byte 1
- sign „-” or space
Byte 2
- space
Byte 3
4 - digit or space
Byte 5÷9 - digit, decimal point or space
Byte 10
- digit
Byte 11
- space
Byte 12
- k, l, c, p or space
Byte 13
- g, b, t, c or %
Byte 14
- space
Byte 15
- CR
Byte 16
- LF