Preparing balance to work
1.
Take the balance, the feeder and mechanical elements of the pan out. It is
recommended to keep the original scale package in order to transport the
balance safely in future.
2.
Place the balance on a stable ground not affected by mechanical vibrations
and airflows.
3.
Level the balance with
rotating legs
6
so that the
air bubble in water level
7
at the back of the
balance is in the middle.
4.
(for
AG100-AG600)
Gently insert the mandrel
of pan support
2
into
balance
mechanism
socket through the pan
ring 3 and the pan
1
on
(AG600 balances have
not the pan ring).
5.
(for AG1000-AG4000)
Place nuts
2
on mandrels
that
are
visible
in
balance cover holes, put
the pan
1
on nuts.
6. If the balance was taken from a lower temperature surrounding to
a room with higher temperature, e.g. in winter, moisture can liquefy on
the balance casing. Do not connect power supply to the balance, because
this can cause damage or improper work of the balance. In this case leave
the balance for at least 4 hours unplugged for acclimatization.
7
6
1
2
3
7
6
1
2