USER MANUAL
43
____________________________________________________________________________________________________________________
14.17 Function of comparing with preset threshold values
This function allows to compare the weighing result with two, previously
programmed values: upper and lower threshold. The comparison result is signalled
with the messages
MIN
,
OK
or
MAX
on the display.
If the weighing result is:
- smaller than lower threshold – the balance signals
MIN
,
- between thresholds – the balance signals
OK
and emits acoustic signal,
- bigger than the upper threshold – the balance signals
MAX.
- smaller than zero threshold (no load) – none of above messages appears.
Press
MENU
key to enter the user
functions
menu,
select
Threshold
signalling
with cursor and press
ENTER
.
Using
∨
and
∧
keys, move cursor to
Zero threshold
option and press
ENTER
.
Enter the indications value, below
which the balance is considered
unloaded (no signalisation) and press
ENTER
.
In the same way enter the values for
Min threshold
and
Max threshold
.
The
Outputs mode
option is used for
setting the THRESHOLDS connection
operation mode (see below). The
appropriate mode is selected with < and
> keys, confirmation – with
ENTER
key.
USER FUNCTIONS&OPT.
Autotaring
PCS counting
. . .
Streshold signalisation
USER FUNCTIONS \ TRESHOLD SIGNAL.
1. Activate
2. Zero treshold : none
3. Min treshold : none
4. Max treshold : none
5. Outputs mode : none
6. Buzzer
: none
7. LCD flashing : OFF
8. Exit
USER FUNCTIONS \ TRESHOLD SIGNAL.
1. Activate
2. Zero treshold : 5
3. Min treshold : none
4. Max treshold : none
5. Outputs mode : none
6. Buzzer
: none
7. LCD flashing : OFF
8. Exit
USER FUNCTIONS \ TRESHOLD SIGNAL.
1. Activate
2. Zero treshold : 5g
3. Min treshold : 90g
4. Max treshold : 110g
5. Outputs mode : <signalling>
6. Buzzer
: stable OK
7. LCD flashing : OFF
8. Exit