14.21.1 Solid body density determination
Suspend the trays and tare scale
pressing button
T
.
Press button
Menu
, select
Density
measurement
with cursor and press
ENTER
..
Select
Material type
with cursor and
press
ENTER
.. Use > and < keys to
choose
solid
material and press
ENTER
Select
Liquid type
with cursor and press
ENTER
.
When is used distilled water please use
> and < keys and select
water
. If is
used a different liquid - please choose
different
, and press
ENTER.
If is used distilled water please write
current temperature of water. This is
necessary to calculate a correct value of
density – require accuracy is 0,5 degree.
If is used a different liquid please write
accurate density value this liquid on the
present temperature.
Finish the procedure pressing
ENTER
.
USER FUNCTIONS
Autotaring
PCS counting
. . .
Density measurement
. . .
USER FUNCTIONS / DENSITY
1.Activate
2.Material type :
solid
3.Liquid type
: water 20.0
o
C
4.Liquid density : 0.99820g/cm3
5.Report printout
6.Exit
USER FUNCTIONS / DENSITY
1.Activate
2.Material type :
solid
3.Liquid type
: water 20.0
o
C
4.Liquid density : 0.99820g/cm3
5.Report printout
6.Exit
USER FUNCTIONS / DENSITY
1.Activate
2.Material type :
solid
3.Liquid type
: < water >
4.Liquid density : 0.99820g/cm3
5.Report printout
6.Exit
USER FUNCTIONS / DENSITY
1.Activate
2.Material type :
solid
3.Liquid type
: water T= 20.0
4.Liquid density : 0.99820g/cm3
5.Report printout
6.Exit