USER MANUAL
27
Calibration options (internal and external):
Except of
Report printout
, calibration options are available after
changing position of adjustment switch.
Press
MENU
key, choose
Calibration
option with the cursor and press
ENTER
.
External load
option allows entering
the value of external weight used for
calibration. Choose
External weight
option with the cursor, press
ENTER
key and use
>
and
<
keys to select
desired value. Several standard values
are available, but it is advised to use
as great weight value as possible.
Time of calibration
option allows
defining the time, after which the
balance performs internal calibration
automatically.
Temp. of calibration
option allows to define the change of
environment temperature, at which
internal calibration will be performed
automatically. Values of the options
are selected as with external load
above.
If external printer is connected to the
balance,
Report printout
option allows
obtaining calibration data of the
balance, which proofs that calibration
process was performed correctly and
may be useful for balance diagnostics.
USER FUNCTIONS&OPT.
Autotaring
PCS counting
. . .
Calibration
. . .
USER FUNCTIONS&OPT. \ CALIBRATION
1. External calibration
2. External load
: 200g
3. Time of calibration
: 2h
4. Temp. of calibration : 2.0
o
C
5. Report printout
6. Exit
USER FUNCTIONS&OPT. \ CALIBRATION
1. External calibration
2. External load
: 200g
3. Time of calibration
: 2h
4. Temp. of calibration : 2.0
o
C
5. Report printout
6. Exit
USER FUNCTIONS&OPT. \ CALIBRATION
1. External calibration
2. External load
: 200g
3. Time of calibration
: 2h
4. Temp. of calibration : 2.0
o
C
5. Report printout
6. Exit