Calibration
Calibration should be carried out after
installation and at regular intervals. When
the software detects an uncalibrated reader
the following dialogue box will appear
Reader XXXX
(serial number)
has not been
calibrated
. You must click OK to continue.
You will also need your calibration card which
should be located in your documentation
wallet.
Use either of the large calibration codes at
the bottom of the card.
Calibration is a quick and simple process.
First open the Axicon linear verifier software
(main program), select
Calibrate
from the
Options
menu of the command window
(or press F2). You will see a window with
two edit boxes for minimum and maximum
reflectance values. Ensure that the values
shown in the window match the values
printed at the bottom of the calibration card.
Modify the figures on the window if necessary
to ensure that they are the same as the card.
With the
Calibration
window open scan the calibration barcode 10 times. Then
click the OK button. The verifier is now calibrated and ready to use. After
calibration, store the calibration card in the documentation wallet, away from
light and physical damage. If your calibration card should ever become damaged or
discoloured, please contact Axicon or your supplier for a replacement.
Axicon 6525-S, 7025-S