R
B
can be measured directly on the test jig (with the test jig disconnected) to determine
calibrated relationships, if desired. R
C
is provided by the user, and can be calibrated as
required.
INCORRECT ORIENTATION
The instrument and the DUT will not be damaged if the diode is installed with the
incorrect orientation (i.e., with the anode and cathode reversed). However, incorrect
waveforms will be generated,
CABLE LENGTHS
The cable lengths are not critical. They may be increased or decreased as desired.
The cables connecting the OUT and MON signals to the oscilloscope should have
identical lengths, to avoid introducing timing skews.
ACCESSIBLE VOLTAGES
The mainframe provides pulsed voltages of up to 100V to the test jig. For this reason,
the output is automatically disabled when the test jig lid is open. The lid must be closed
to obtain measurements.
Shielded cabling should be used for all connections to the "IN" and "OUT"
terminals on the test jig, and the "OUT" connector on the mainframe.
When used properly (with R
C
= 50 Ohms), the maximum voltage on the OUT
terminal will be < 1V, approximately. However, if R
C
is not connected and the DUT is
not installed, the maximum voltage will at the OUT terminal may be as high as 100V.
You may need to consider whether this scenario would damage your oscilloscope, and
take appropriate precautions.
27
Summary of Contents for 14260
Page 5: ...Last modified July 28 2022 Copyright 2022 Avtech Electrosystems Ltd All Rights Reserved 5...
Page 46: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY...
Page 47: ...PCB 235D HIGH VOLTAGE DC POWER SUPPLY...
Page 48: ...PCB 104H KEYPAD DISPLAY BOARD...
Page 49: ...PCB 201C INTERLOCK CONTROL...
Page 50: ...MAIN WIRING...
Page 52: ...TEST JIG WIRING AVX TFR SOD123W...
Page 53: ...TEST JIG WIRING AVX TFR SOD128...
Page 54: ...PERFORMANCE CHECK SHEET 54...