AWT35-501813
Issue AA
Page 18
Factory default: SPEEd 1 (standard 15Hz)
LF 5
Zero Tracking
: use this to define a ± 0 /5 divisions range around zero. When scale
weight is not at the center of zero but inside this range, ½ of the weight will be subtracted
until the weight is inside the center of zero region.
Factory default: ZP 0 (off)
LF 6
Factory default: none
LF 7
Gravitational value
:
use this item to key in a G constant value: 9.78031 < G < 9.83217
If the scale has been calibrated at a different location and it is not possible to re-calibrate
with known test weights, the scale can be adjusted using this gravity factor.
•
Current G value can only be edited if Cal Switch ON. See chapter 6. Calibration.
Factory default: 9.81259
CAUTION!
Verify with local agencies if adjusting the gravity factor is accepted in your
area. It may be required that calibration be done with certified weights.
LF 8
Zeroing
at Every Power UP:
use this items to set an auto zero at the power up.
Factory default: SEtZ Y (on)
4.3 Quick Calibration Menu
Before entering this menu, make sure that UF-5 is set as HOLD 0.
See chapter 6. Calibration on page
21 for a detailed procedure.
4.3.1. Quick Calibration Menu Levels [ECF]
Navigate the options by pressing
ZERO
(◄),
TARE
(►) and increase a numeric value by
SELECT
(▲)
key. Press
UNITS
(
ESC
) to exit the option or the menu and
(
↵
)
key to confirm.