4.3.2 Quick Calibration Functions Descriptions [ECF]
There are 3 Calibration Functions.
ECF-1
Calibration
: ZERO / SPAN. Test weight should be at least 1/3 of the max. capacity
ECF-2
Zero Calibration
(only)
ECF-3
Span Calibration
(only)
5. Scale Configuration [LF 2]
Enter the function
LF 2
to set the units, counting mode, calibration units, decimal position of the
separator and division size.
Navigate the options by pressing
ZERO
(◄),
TARE
(►) and increase a numeric value by
SELECT
(▲)
key. Press
UNITS
(
ESC
) to exit the option or the menu and
(
↵
)
key to confirm.