36
Medium priority audio alarms are: 3 beeps and 25-second pause. This sequence
repeats until the alarm is cleared or silenced.
Tap on “PAUSE AUDIO” button to pause the alarm audio for 2 minutes. Tap on
“AUDIO OFF” button to permanently turn off the alarm audio.
See “Alarm Limits” section for adjusting the alarm limit.
NOTE:
The following table describes alarm conditions and visual indicators.
Alarm Condition
(Medium Priority Alarm)
Visual Indicator
Oximeter Sensor Probe Detached from the
Finger
(1)
Alarm text message in yellow
blinks rapidly on the screen.
(2)
Alarm LED indicator on the
Oximeter Module blinks
yellow.
Oximeter Sensor Cable/Connection Cable
Detached
Oximeter Module Battery Low
Oximeter Module Detection Error
Display Unit Battery Low
(1)
Alarm text message in yellow
blinks rapidly on the screen.
(2)
Relevant icon highlighted in
yellow blinks rapidly on the
screen.
Thermometer Module Battery Low
Oximeter Module Data Update Period Exceeds
Limit (More Than Half A Minute)
Aulisa sensor module(s) BLE Disconnected
Receiver/Transponder Wi-Fi Disconnected