![Aulisa Guardian Angel Rx Guardian Angel Rx Lite GA2000 Series Instructions For Use Manual Download Page 35](http://html1.mh-extra.com/html/aulisa/guardian-angel-rx-guardian-angel-rx-lite-ga2000-series/guardian-angel-rx-guardian-angel-rx-lite-ga2000-series_instructions-for-use-manual_3016359035.webp)
34
Alarms and Limits
Alarm Features
The GA2000 Series provides high and medium priority audible and visual alarms. The
visual alarm is indicated by the alarm window on the screen. Audio alarms will sound
from the device speaker.
High Priority Alarms
High priority alarms are those that require immediate attention to the person
being monitored, including SpO
2
, pulse rate, and temperature alarms. On the
screen, high priority alarms are indicated with rapid blinking vital sign readings in
red color and with alarm text message when alarm limits are met or exceeded.
(See figure below.)
NOTE:
Alarm LED indicator on the Oximeter Module will blink red along with displays
on the screen.
NOTE:
The volume for high priority alarms cannot be adjusted on the Display Unit.