![Aulisa Guardian Angel Rx Guardian Angel Rx Lite GA2000 Series Instructions For Use Manual Download Page 32](http://html1.mh-extra.com/html/aulisa/guardian-angel-rx-guardian-angel-rx-lite-ga2000-series/guardian-angel-rx-guardian-angel-rx-lite-ga2000-series_instructions-for-use-manual_3016359032.webp)
31
Device Verification
Verify the device function
Step 1:
Set up the system. Refer to the Aulisa sensor module(s) Instructions for Use
for setting up instructions.
Step 2:
Make sure Aulisa sensor module(s) is worn on or attached to the right place
firmly.
Step 3:
Verify that the Bluetooth connection status icon on the screen is blue and
the status indicator on Aulisa sensor module(s) is blinking green.
Step 4:
Verify that the customer Wi-Fi network connection is stable.
Step 5:
Verify that the vital sign readings are displayed.
Step 6:
Verify that the live video is displayed on the MAIN screen.
Verify the alarm function BEFORE each use
Step 1:
Set up the system. Refer to the Aulisa sensor module(s) Instructions for Use
for setting up instructions.
Step 2:
Ensure the Bluetooth and Wi-Fi connections are established, respectively.
(See “Device Connection” and “Device Pairing” sections.)
Step 3:
Set the alarm limit lower or higher than the displayed value.
Step 4:
Verify that an alarm message is displayed and that an alarm audio is
generated. (See “Troubleshooting” section if an alarm message and audio
signal is not generated.)
Step 5:
Press the "PAUSE AUDIO" button to temporarily silence for 2 minutes.
Step 6:
After alarm signal is regenerated, press the "AUDIO OFF" button to silence
permanently the alarm signal.
Step 7:
Tap on “RESTORE DEFAULTS” to restore alarm limits to manufacturer
configured values before use.
NOTE:
Check the top right corner of the MAIN screen. Ensure the Receiver/Transponder
connection indicator is lit blue and the Wi-Fi connection indicator has a strong signal.