MAN - PAV4 - REV: Original | D. Installation
9
3. If equipped, connect the optional UPS battery backup supply.
a. Use the supplied standard power cord to plug the UPS into the wall. The system should charge for 24
hours before connecting it to the PAV.
b. Attach the supplied cord from the UPS to the PAV.
c. See manufacturer’s literature for further setup instructions.
4. Check the flow adjust valve on the PAV (marked Flow Control) to ensure it is set at the correct values. This valve
controls the rate at which the PAV depressurizes. When the flow adjust valve is set at the recommended setting, the
PAV will depressurize at optimal rate. The unit will have a sticker (see below) with the correct values. Use the value
marked Flow Adj.
NOTE: See Section I.Maintenance for information on what to do if the flow adjust valve detaches.
D.5 Regulator Connections
1. Secure the air connection to the air tank using the supplied male and female connections.
2. Connect the hose to the regulator.
NOTE: Do NOT connect the hose to the part marked FLOW. This is the pressure relief valve, and
nothing should be connected to it.
Figure D.3: Sticker with values
Figure D.4: Flow Control valve