MAN - PAV4 - REV: Original | C. System Overview
6
RTD Temperature
Measurement
0.1°C Resolution and ± 0.1° accuracy Microprocessor Temperature
Control with overlimit protection
Power Requirements
230/240 VAC, 10 amps, 50/60 HZ
Vessel Per ASME Code
Section VIII
Division 1; 1992 A 93 Contact ATS with the PAV’s serial number for
a copy of the certification
Vessel Design
Conditions
325 psig (2.24 MPa) at 110°C/121°C
Pressure Safety
Release Valve
340 psi
Temperature Uniformity
± 0.5°C
Pressure Display
Accuracy
±1%
“TIME” Timer Accuracy
0.1%
“OUT OF RANGE”
Timer Accuracy
0.1%
Air Inlet
Male 1/4” NPT Fitting
Solenoid Valve
Closed when power is applied. In the event of power loss, solenoid
valve will open and release any pressure in the PAV vessel.
C.4 Environmental Conditions
Operation of the ATS Pressure Aging Vessel involves use of high temperatures and contents being contained at
high pressure. The location and placement of the PAV should always take these elements into consideration to
ensure a safe working environment. The PAV is meant for use in laboratory/factory settings in a dry and clean work
environment. The work surface that the PAV is placed on should be a clean and sturdy work surface at a reasonable
working height and away from any water, gas, or electrical hazards. The work surface should be able to support more
than the weight of the unit itself. The indoor area that the PAV is placed in should be well ventilated and containing no
open flames or materials that may constitute a fire hazard.