MAN - PAV4 - REV: Original | G. Verification
27
G.2 Verifying Pressure
1. Remove the screw in top of the vessel lid.
2. Screw the pressure gage into that space. Make sure the lid is secured according to the directions in Section F, step
14.
3. Apply the pressure to the PAV. Allow pressure to reach 2.1 ± 0.1 MPa and stabilize for fifteen minutes.
4. Verify that the display on the pressure gage matches the displayed pressure on the PAV display.
5. Note any difference and call ATS for calibration if the difference falls out of specification.
NOTE: It is best to have the PAV setpoint at a low temperature.
NOTE: Do not verify temperature and pressure at the same time.
Figure G.5 - Screw on Vessel Lid