MAN - PAV4 - REV: Original | E. Software Overview
20
Stop Test on all Errors Checkbox
Selects what the system will do if an error is generated. If checked the system will stop the test and display the alarm
screen when any of the following errors happen:
• Over temperature
• Recovery Error Limit
• Age Error Limit
• No Pressure Error
• Standardization Error
Current Time
Allows the operator to set the current time of the system. This will affect any log files created.
Preheat Delay
Used to allow the system to be set up to start automatically after a time delay. You can think of it like a “coffee timer”.
Just enter the Hours and Minutes you wish to wait before preheat start and press the yellow button. The button
will light to show that a delay has been set. The delay button is a toggle switch, so to turn off a delay that has been
started just press the button again. The status of the system will show “Preheat Delay” when it is active and the
remaining time before preheat start is shown.
Done Button
The Done button will return you to the View Screen.
E.6 Alarm Screen
Figure E.16 shows the Alarm Screen. It is shown anytime the Alarm button is pressed on the View Screen, also when
the system detects an alarm or warning and the “Stop Test on all Errors” checkbox is checked.
It will show all current alarms (in Red) and warnings (in Yellow). You can try to reset these with the “Reset” button.
Some alarms will go away automatically when the alarm condition goes away.
Alarm Off Button
Will turn off the alarm buzzer if it is on.
Done Button
Will return you to the Main Screen.