User Manual
California Instruments
INITIAL SETUP 3
VOLT SENSE
Sets the power-on state of the voltage sense mode. Available
settings are Internal (INT) or External (EXT).
WAVE GROUP
[] Sets the user defined waveform group that will be loaded at
power on. Available groups are 0, 1, 2 and 3. Each group can
contain up to 50 user-defined waveforms. A waveform group
can only be loaded at power up. To change groups, you must
change this field to the desired new group and cycle the power
to the AC source or issue a *RST command over one bus.
CLOCK MODE
Sets the clock source used at power up. Available settings are
Stand Alone (STAND), MASTER, and SLAVE.
NO. OUTPUT
[] Sets the phase mode at power up. Available options are
determined by the presence of the phase mode option. If the
phase mode option is installed, available options are single
phase mode (ONE), three phase mode (THREE) or last used
phase mode (LAST). If the phase mode option is not installed,
this field is either fiXed to ONE or THREE.
VOLT ALC
Determines ALC mode at power on. The ALC mode adjusts the
output voltage based on internal voltage measurement system
and provides enhanced output regulation and accuracy.
Available settings are ON, OFF or REG.
98
i Series II / NSG 1007 II / iM Series II
Summary of Contents for 10002iX
Page 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Page 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Page 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Page 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Page 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...