4.2.11.3INITIAL SETUP menu
Figure 4-51: INITIAL SETUP menus
Any time the AC source is powered up, the output will reflect the values stored as the INITIAL
setup values. This allows the unit to be powered up in a known state at all times. The INITIAL
values can be set in the INITIAL SETUP menus.
The initial setup can be used to power up the AC source with the output on and a high voltage
present at the output. For normal situations, this is not recommended due to the potential danger
to the operator. It is recommended that the initial voltage be set low and/or the output relay be
programmed to OFF for most situations.
The following fields are provided in the INITIAL SETUP menus:
Entry
Description
INITIAL SETUP 1
VOLTAGE
Sets the power-on AC voltage for AC and AC+DC modes or the
DC voltage for DC mode.
CURR LIMIT
Sets the power-on current limit value.
FREQ
Sets the power-on frequency value.
PHASE
Sets the power-on frequency for phase A with respect to an
external sync signal. If the internal oscillator is used (default)
this setting has no effect.
INITIAL SETUP 2
VOLT RANGE
Sets the power-on voltage range value. The available choices
are determined by the VOLT PAIR selected in the
VOLTAGE/CURRENT CONTROL SETUP menu.
VOLT MODE
Sets the power-on voltage mode. Available settings are AC
mode, DC mode or AC+DC mode [].
OL MODE
Sets the power-on overload mode. Available settings are
Constant Current (CC) or Constant Voltage (CV) mode.
OUTPUT RELAY
Sets the power-on state of the output relay. Available settings
are ON or OFF.
NSG 1007 Series II
97
Summary of Contents for 10002iX
Page 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Page 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Page 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Page 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Page 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...