User Manual
California Instruments
4.4Waveform Management
[]
The NSG 1007 employs independent arbitrary waveform generators for each phase. This allows
the user to create custom waveforms. In addition, the iX offers three standard waveforms that
are always available. This chapter covers issues that relate to defining, downloading and
managing custom waveforms.
4.4.1Standard Waveforms
For many AC applications, a sine wave shape is used. The sine wave is one of the standard
waveforms provided on the NSG 1007. The standard sine wave is always available and is the
default waveform at power-on. In addition to the sine wave, two more standard waveforms are
available, square and clipped.
Figure 4-57: Selecting a waveform
The square wave provides a high frequency content waveform with relative fast rise and fall
times. Due to AC amplifier bandwidth limitations, the frequency content of the standard square
wave has been kept within the amplifier’s capabilities. As the fundamental frequency is
increased, the relative contribution of higher harmonics is reduced.
The clipped sine wave may be used to simulate voltage distortion levels to the unit under test.
The total harmonic distortion level may be programmed in percent using the CLIP LEVEL field of
the WAVEFORMS menu. Changing the distortion level of the CLIP waveform forces the AC
source to regenerate the CLIPPED sine wave’s data points and reload the waveform register with
the newly requested data. This process requires the output to be dropped briefly. To avoid
interrupting the voltage output to the unit under test, select a different waveform such as the
standard sine wave first, change the clip level and change the waveform back to the CLIPPED
sine wave. This will avoid any output interruption.
4.4.2Phase Selection
Figure 4-58: Selecting waveforms for single phase or all phases
If the NSG 1007-9 orNSG 1007-15-208, different waveforms may be selected for each phase.
The number of custom waveforms from which to select remains 50 but each phase can be
assigned a different custom or standard waveform. The specific output phase for which the wave
shape is programmed is selected with the PHASE key on the front panel. The selected phase is
always shown in the top right hand corner of the WAVEFORMS display.
To select the same wave shape for all three phases in a three phase configuration, press the
PHASE key until the øABC enunciator appears in the top right corner of the WAVEFORMS
menu. Waveform selections made in this mode will apply to all three phases.
NSG 1007 Series II
107
Summary of Contents for 10002iX
Page 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Page 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Page 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Page 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Page 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...