User Manual
California Instruments
Parameter
Range
Accuracy (
±
)
Resolution
Accuracy specifications apply above 100 counts. Current and Power Accuracy specifications are times two for NSG
1007-10-208 and times three for NSG 1007-15S-208. For NSG 1007-10-208 and NSG 1007-15S-208, resolution
decreases by factor of 10, ranges for current and power increases by factor of three.
2.1.5Harmonic Measurements (NSG 1007)
Parameter
Range
Accuracy (
±
)
Resolution
Frequency fundamental
16.00 - 1000 Hz
2 counts
0.01 Hz to 1 Hz
Frequency harmonics
32.00 Hz - 16 kHz
2
°
typ.
0.5
°
Voltage
Fundamental
0.25V
0.01V
Harmonic 2 - 50
0.25V + 0.1% + 0.1%/kHz
0.01V
Current
Fundamental
0.05A
0.01A
Harmonic 2 - 50
0.05A + 0.1% + 0.1%/kHz
0.01A
Accuracy specifications are times three for three phase mode. Harmonics frequency range in three-phase mode is
32 Hz - 16 kHz. Resolution decreases by factor of 10 for NSG 1007-10-208 and NSG 1007-15S-208.
2.1.6System Specification
Parameter
Specification
External Modulation:
0 to 10%
Synchronization
Input:
Isolated TTL input for external frequency control. Requires 5V at 5 ma
for logic high.
Trigger Output:
400
µ
s pulse for voltage or frequency change. Isolated output that
requires a pull-up resistor, 22K
Ω
, to + 5 VDC.
Non volatile memory
storage:
16 complete instrument setups and transient lists, 100 events per list.
Waveforms
Sine (i series)
Sine, square, clipped, user defined (NSG 1007)
Transients
Voltage: drop, step, sag, surge, sweep
(i/iX only)
Frequency: step, sag, surge, sweep
Voltage and Frequency: step, sweep
IEEE-488 Interface:
SH1, AH1, T6, L3, SR1, RL2, DC1, DT1
IEEE 488.2 and SCPI
Response time is 10 ms (typical)
RS232C Interface:
Bi-directional serial interface
9 pin D-shell connector
Handshake: CTS, RTS
Data bits: 7, 8
Stop bits: 1,2
Baud rate: 9600, 19200, 38400 (Models without USB I/F)
Baud rate: 9600, 19200, 38400, 57600, 115200, 230400, 460800
IEEE 488.2 and SCPI.
28
i Series II / NSG 1007 II / iM Series II
Summary of Contents for 10002iX
Page 4: ...User Manual Teseq 4 NSG 1007 Series II ...
Page 48: ...User Manual California Instruments 48 i Series II NSG 1007 II iM Series II ...
Page 53: ...User Manual California Instruments Figure 3 20 Functional Test Setup NSG 1007 Series II 53 ...
Page 135: ...User Manual California Instruments Figure 5 75 Logic Board LED s NSG 1007 Series II 135 ...
Page 205: ...User Manual California Instruments Figure 9 115 NSG 1007 15 208 3 EOS3 NSG 1007 Series II 205 ...