TD 92816EN
17-July 2012 / Ver. B
39
Programming and Installation Manual
Care Phone
Repeat sequence
In this menu, specify how many times the sequences shall run. The value can be
set between 1 and 9, the factory default is 1. Note that the value applies to all
four sequences. Max 40 attempts in total is possible.
Demand disconnection
In this menu choose what Analog Care Phone / GSM Care Phone should do if the
call does not disconnect properly, but speech communication has been
established.
Choose <Yes> Analog Care Phone / GSM Care Phone will call again.
Choose <No> Analog Care Phone / GSM Care Phone will not call again. The
default value is <No>.
Note:
If you receive alerts in STT Mobile alarm, the above set-up requiring disconnection
will not be valid. There the alarm is disconnected even if speech communication has been
disconnected as a member of staff has opted to take it, and therefore has the
responsibility for this alarm.
9.4
Attempts
Has a submenu where you can specify how many attempts Analog Care Phone /
GSM Care Phone shall make to contact the alarm receiver on the respective
phone number/address.
>Number A< to >Number J< >Nightmode N< >Temp redir. V< >Address a< to >
Address h< >Nightmode n< >Temp redir. v<
You can enter values between 1 and 9. If you enter values greater than 9, Analog
Care Phone / GSM Care Phone will still not make more than 9 attempts. The
default value is 5. Remember that an attempt may take from several minutes
(depending on whether the Care Phone reaches the alarm receiver or not) so
there should not be too many attempts on the primary number.
9.5
Connection type
This menu is only available in Careline GSM and has a submenu where for each
phone number and address you can specify the type of connection used.
>Number A< to >Number J< >Nightmode N< >Temp redir. V< >Address a< to >
Address h< >Nightmode n< >Temp redir. v<
On the phone numbers you can specify whether the alarm is to call via the <Tele
line> or <Wireless>. For the addresses, you can in the Careline GSM only enter
<Wireless>, as this communication goes over GPRS.
9.6
Outgoing call
Has a submenu, which controls functions that are used when the Care Phone is
dialling.
>Precall signal< >Dial tone detection< >Mute dialling< Polarity change<