Technical Specifications
Synergy
RF™
User’s Guide
DFU-0221-6r0_fmt_en-US
Page 22 of 58
3.5
Safety, EMC, and Regulatory Requirements
Table 10
Safety, EMC, and Regulatory Requirements
Parameter
Parameter Value
System
Classification
IEC 60601-1
Class I
(protection against electric shock)
FDA Class
Class II
EU Class
Class IIb
Health Canada Class
Class 3
Safety
Certifications
Domestic Certification
ES60601-1:2005/A1:2012
Canadian Certification
CAN/CSA C22.2 No. 60601-1-08
EU Certification
EN 60601-1:2006/A1:2013
EMC
Certifications
CISPR 11 EMC Class
Class A
CISPR 11 EMC Group
Group 1
EMC Certification
Certification to EN 60601-1-2:2015\
IEC 60601-1-2:2014, 4
th
edition
Safety
Certification
Marking
CE
Classification
93/42/EEC
Annex IX Rule 9
Class IIb
For all other accessories refer to accompanying DFUs for more information.
Refer to Section 13.0 Electromagnetic Emissions for further details on EMC
Certification