Operation and Frequently Used Functions
Synergy
RF™
User’s Guide
DFU-0221-6r0_fmt_en-US
Page 42 of 58
5.2.5
Footswitch Override Screen
The
Footswitch Override
screen has radio buttons to select the footswitch
override mode for each channel.
At the End User’s discretion, the default setting can be changed so that the
Arthrex Apollo
RF®
device (Probe) can be controlled by both the Footswitch and/or
the Hand Controls, as described below. If both the Footswitch and the Hand
Controls are enabled and activated simultaneously, whichever button is
activated first will take precedence and other inputs will be ignored while that
button is activated.
W A R N I N G !
When the Footswitch Override function is disabled the disposable RF
device (Probe) hand controls are active. Care must be taken not to activate
the device inadvertently.
In the default mode, the buttons on the hand-control of the Arthrex Apollo
RF®
probe will not function when a footswitch is attached to the console. The End
User may select the both footswitch and hand controls option and the buttons on
Summary of Contents for Synergy AR-9800
Page 2: ...This page intentionally left blank ...
Page 26: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 24 of 58 ...
Page 27: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 25 of 58 ...
Page 28: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 26 of 58 ...
Page 29: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 27 of 58 ...
Page 30: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 28 of 58 ...