Product Description and Intended Use
Synergy
RF™
User’s Guide
DFU-0221-6r0_fmt_en-US
Page 20 of 58
Message
Explanation
"Power Supply PFC Voltage"
Power supply self-check failure, contact
Arthrex technical services.
"Power Supply Excessive Voltage"
Power supply is exceeding the voltage
parameters, contact Arthrex technical services.
"Power Supply Excessive Power"
Power supply is exceeding the power
parameters, contact Arthrex technical services.
"Relay Failure"
Power supply self-check failure, contact
Arthrex technical services.
"RF Output Failure"
Power supply self-check failure, contact
Arthrex technical services.
"Touchscreen Circuit Failure"
Touch panel failure, contact Arthrex technical
services.
Figure 4
Error Screen Display
Table 5
Error Screen Display – Descriptions
1.
Error Message
The description of the error/fault experienced in the
system
2.
Error Code
The
Error Code
specifies which error occurred for
additional reference for Arthrex technical services
3.
Explanation
Provides context of the error/fault
1
2
3
Summary of Contents for Synergy AR-9800
Page 2: ...This page intentionally left blank ...
Page 26: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 24 of 58 ...
Page 27: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 25 of 58 ...
Page 28: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 26 of 58 ...
Page 29: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 27 of 58 ...
Page 30: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 28 of 58 ...