DrillSaw Sports 400™ System
User’s Guide
DFU-0253-4r0_fmt_en-US
Page 13 of
46
3
3
.
.
0
0
Technical Specifications
Handpiece
Table 5
Handpiece AR-400
Width
40 mm (1.574”)
Height
141.5 mm (5.57”)
Depth
121mm (4.764”)
Weight
630 g (1.3 lbs.)
Cannulation
3.2 mm (0.126”)
Protection
IPX0 for Handpiece and IPX4 for AR-400UBH-1
Speed
See Section 3.2 Table 7 Attachments
Torque
See Section 3.2 Table 7 Attachments
Applied part type
B
Pollution level
2
Overvoltage category II
Altitude
Up to 3,000 m [9,842 ft.] above sea level
Duty Cycle*
Operating Type S3 per IEC 60034-1
At 25°C (77° F) Ambient Temperature:
30 seconds on/off for 5 cycles, then wait 48 minutes
Worst Case Temperature Condition: At 40°C (104°F)
18 seconds on/off for 5 cycles, then wait 60 minutes
*
Following the duty cycle is an important part of maintaining the Arthrex DrillSaw
Sports 400 Power System. The duty cycle times have been determined under average
load, using the Sagittal Saw attachment and at ambient air temperature of 25° C (77°
F). Operating the device above of these prescribed parameters can lead to user and
patient injury, as well as a shortened life span of the device itself. The recommended
environmental operating temperature is 25° C (77° F) or lower. Operating in
environments above 25° C (77° F) can shorten the above prescribed duty cycles. The
environment temperature, surgical attachment type and time need to be considered
by the user, when planning its use in surgical interventions. Additionally, please
read the
General Warnings and Safety Notices
that start on Page 1 of this User’s Guide.
Table 6
Ambient conditions for operation
Temperature
10° to 40 °C (50° to 104 °F)
Relative Humidity
15% to 80% (relative), non-condensing
Table 7
Ambient conditions for storage (in shipping packaging)
Temperature
-40° to +70°C (-40° to +158°F)
Relative Humidity
8% to 80% (relative), non-condensing