DrillSaw Sports 400™ System
User’s Guide
DFU-0253-4r0_fmt_en-US
Page 27 of
46
Aseptic Battery Housing/Aseptic Battery Pack/Aseptic
Transfer Kit
These instructions describe the procedure for transferring the aseptic battery pack
into the aseptic battery housing, using the aseptic transfer kit.
Figure 15
Aseptic Battery Pack Transfer
Table 20
Aseptic Battery Pack Transfer
A
Sterile Technician
B
Non-sterile Circulating Nurse
1.
The Sterile Technician holds the aseptic battery housing (sterile)
upright and places the aseptic transfer kit (sterile) on top.
2.
The Non-sterile Circulating Nurse places the aseptic battery pack (non-
sterile) down into the aseptic battery housing (sterile) through the
aseptic transfer kit (sterile) until the aseptic battery pack is fully seated.
3.
The Non-sterile Circulating Nurse then removes the aseptic
transfer kit, and the Sterile Technician closes the sterile battery
housing door.
1
2
3
A
A
B
A
B
3a