Cleaning and Disinfecting
Synergy Resection
User’s Guide
DFU-0213-7 Rev. 0
Page 38 of 50
Remove any accessories in the device. This includes shaver blades, saw blades or
drill bits.
If the handpiece has suction control, make sure it is fully open.
Clean the handpiece with an enzymatic cleaner without subsequent acid
neutralization.
If the handpiece has suction control, clean the connection with a cleaning brush.
Rinse the handpiece thoroughly after cleaning.
Disinfect the handpiece with a commercially available disinfectant.
Thoroughly rinse the handpiece under lukewarm distilled water or preferably
demineralized sterile water.
See Section 7.0 for sterilization information.
Always comply with the instructions issued by the manufacturer of
the disinfectant.
NEVER place the shaver handpiece in Cidex or other aldehyde
disinfectant solutions.
NEVER allow the console plug pins to have any contact with
liquids. Remove dust or moisture with dry compressed air. Only
DRY connectors may be plugged into the console.