General Warnings and Safety Notices - Read This First
Synergy Resection
User’s Guide
DFU-0213-7 Rev. 0
Page 1 of 50
1.0 General Warnings and Safety Notices - Read This First
Important Symbols and Conventions
It is imperative that the symbols and conventions listed below be clearly
understood. The
Synergy Resection User’s Guide
identifies critical, important, and
useful information using these symbols and conventions.
Users of this device are encouraged to contact their Arthrex representatives if
they require a more comprehensive surgical technique.
W A R N I N G !
The WARNING! is the most important safety symbol. It identifies
critical
information that must be followed precisely to avoid injury or death.
1. Use this device only for purposes described in the User’s Guide, under the
supervision of a trained and licensed physician.
2. No modifications of the console (AR-8305) or accessories are allowed.
3. Use this device only under the supervision of a trained and licensed
physician. This device should not be used by untrained personnel or used for
indications other than those described in this User’s Guide.
4. Do not open or attempt to service this system, as this may void your
warranty. There are no user-serviceable parts inside. Removing the
cover may introduce an electric shock hazard by exposing you to
dangerous high voltages or other risks. If the system malfunctions,
return it for service immediately.
5. To avoid the RISK of electric shock, this equipment must only be connected
to a SUPPLY MAIN with protective earth ground.
6. Do not have device in direct contact with patient if high-frequency devices are
in use, or if the patient requires defibrillation.
7. DO NOT stack or place equipment adjacent to the AR-8305 console, if
possible. If such a configuration is necessary, carefully observe the
configuration in question to ensure that electromagnetic interference does not
degrade performance.
8. Use only Arthrex approved accessories. Other accessories may result in
increased emissions or decreased immunity of the system. Contact your
Arthrex representative for a complete list of accessories.
DO NOT
modify any
accessory. Failure to comply may result in patient and/or operating room staff
injury.
9. This equipment is
NOT
suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen rich or nitrous oxide environment.
10. Do not have device in direct contact with patient if high-frequency devices are
in use or if the patient requires defibrillation.
11. After autoclaving, the accessory devices are VERY HOT. Handle with care to
avoid burns.