General Warnings and Safety Notices - Read This First
Synergy Resection
User’s Guide
DFU-0213-7 Rev. 0
Page 2 of 50
The PRECAUTION! symbol identifies methods and procedures
that must be followed to avoid damaging the device or causing it
to malfunction.
1. DO NOT
– under any condition or for any reason – open the console (AR-
8305) or any other shaver accessory.
2. Only use
replacement power cords that meet medical grade standards
according to IEC 60320-1, Detachable Power Supply Cords or electrical
standards for the designated country where the AR-8305 is being used.
Contact your Arthrex representative for further information.
3.
Avoid
position console so that it is difficult to disconnect coupler or plug from
supply main.
4.
C
onnecting an extension cord to the AR-8305 may result in a reduced level of
safety.
5. Always
use fuses with the correct values to avoid allowing overcurrent to
enter the system.
6.
An incorrect fuse may increase the risk of electrical shock or fire hazard.
7.
This device has passed testing for EMI / RFI radiation and susceptibility and
EMC compatibility. This device may cause interference to other devices in the
near vicinity if not set up and used as Arthrex instructs.
8. Do not
attach handpieces or footswitches during the Self Test or the
Programming Modes.
9. Only use
footswitches developed by Arthrex specifically for the AR-8305
Synergy Resection console.
10. Only use
shaver handpieces and cables that have been developed by
Arthrex specifically for the Synergy Resection console.
11. Only use
handpieces and cables that have been developed by Arthrex
specifically for the Synergy Resection. The footswitch cable connects and
locks to the console to prevent accidental separation during use. To avoid
damage, disconnect the footswitch by pulling on the cable connector shell
(plug) only.
12.
The accessory handpiece cable connects and locks to the console to prevent
accidental separation during use. To avoid damage, disconnect the
footswitch by pulling on the cable connector shell (plug) only.
13. Always
comply with the instructions issued by the manufacturer of the
cleaning disinfectant regarding concentration, exposure times,
temperature and material compatibility
.
14. NEVER
clean the console with liquid cleaners. If there is dust, remove it with
dry compressed air.
15. Do NOT clean the device with abrasive cleaning or disinfectant
compounds, solvents, or other materials that could scratch or damage
the device.
16. The footswitch is NOT suitable to be cleaned and disinfected in a
thermo washer disinfector.
17. NEVER
allow the console receptacles to have any contact with liquids. If
there is dust or moisture on the receptacles, remove it with dry compressed
air. ONLY dry connectors should be plugged into the console.