Electromagnetic Emissions
Synergy Resection
User’s Guide
DFU-0213-7 Rev. 0
Page 48 of 50
Immunity
test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.7
GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the Model
AR-8305, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
[
]
P
V
d
1
/
5
.
3
=
= 1.17
P
[
]
P
V
d
1
/
5
.
3
=
=1.17
P
80 MHz to 800 MHz
[
]
P
E
d
1
/
7
=
= 2.33
P
800 MHz to 2.7 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the Model AR 8305is used exceeds the applicable RF compliance level
above, the Model AR-8305 should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the Model AR-8305
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.