Sterilization
Synergy Resection
User’s Guide
DFU-0213-7 Rev. 0
Page 39 of 50
7.0 Sterilization
Sterilization capabilities, cleaning, disinfecting, handling, and storage of instrumentation
are the responsibility of qualified facility and/or user personnel. Qualified personnel
must still properly clean and disinfect the instruments prior to sterilization.
Refer to the Instructions for Use package insert for detailed Shaver
Handpiece (DFU-0154-XX or DFU-0240-XX) and Adapteur Power
System II Accessory Attachments (DFU-0149-XX) cleaning and
sterilization instructions included with each handpiece. Additional
copies of this insert can be obtained from the Arthrex website at
www.Arthrex.com, or by contacting your local Arthrex
Representative.
After sterilization in the autoclave, let the accessory device cool
down slowly. NEVER use cold water to cool the handpieces. This
will damage the electronic components and seals.
Liquid on the cable connector of the accessory device can damage
the device. Before connecting the cable, ensure the receptacles
are clean and dry.
W A R N I N G !
After autoclaving, the accessory devices are VERY HOT. Handle with care to avoid burns.
The following special instructions are to be observed when sterilizing
accessory handpieces.
Open or remove the suction control completely to ensure proper
circulation, if applicable.
Open the chuck completely to ensure proper circulation, if
applicable.
Remove all accessories, including shaver blades, drill bits and saw
blades before sterilizing a handpiece.