Specifications
All values relate to a one-hour warm-up period.
Description
5400-15-28 5400-30-28 5400-20-56
Drive Channel
Current
Range (A)
15
30
20
Compliance Voltage (V)
28
28
56
Max Power (W)
420
840
960
Resolution (A)
0.01
0.01
0.01
Accuracy (±[% set point + A])
0.5 + 0.01
0.5 + 0.01
0.5 + 0.01
Noise/Ripple (mA, rms)
< 20
< 30
< 25
Temperature Control
Range (°C)
1
-99
to
250
Resolution (°C)
0.001
2
Thermistor Accuracy (±°C)
3
0.05
4
AD560 Accuracy (±°C)
3
0.05
LM335 Accuracy (±°C)
3
0.05
RTD Accuracy (±°C)
3
0.05
Short Term Stability (1hr) (±°C)
5
0.004
Short Term Stability (24hr) (±°C)
5
0.01
Measurement Channels
Current
Resolution (mA)
10
Accuracy (±[% r mA])
0 + 30
0 + 60
0 + 30
Voltage
Resolution (mV)
10
Accuracy (±[% r V])
0 + 0.05
Sensor
6
10
μ
A Thermistor
Range (k
Ω
)
0.1 – 450
Resolution (k
Ω
) 0.01
Primary Accuracy
(± [% read + k
Ω
])
0.05
+
50
Aux Accuracy
(± [% read+ k
Ω
])
0.20
+
50
100
μ
A Thermistor
Range (k
Ω
) 0.05
–
45
Resolution (k
Ω
) 0.001
Accuracy (± [% r k
Ω
]) 0.05
+
5
1
Software limits. Actual range dependent on sensor type and system dynamics
2
RTD and auxiliary sensor resolution 0.01°C
3
Accuracy figures are the additional error the 5400 adds to the measurement, and does not include the
sensor uncertainties.
4
25°C, 100
μ
A thermistor
5
Stability measurements done at 25°C using a 10k
Ω
thermistor on the 100
μ
A setting. The number is ½
the peak-to-peak deviation from the average over the measurement period.
6
Specifications apply to both primary and auxiliary sensors unless otherwise indicated.