Page 32 · 5400 Series
TECSource
User’s Manual
Out Off Menu
The entries in the
Out Off Menu
are there to control which conditions of the
temperature controller will turn off the output. The following conditions are listed,
along with their defaults:
Thi Limit
(default
Yes
)
Tlo Limit
(default
Yes
)
Sensor Open
(default
Yes
)
Sensor Short
(default
Yes
)
Apply Nonactive
(default
No
)
Apply Aux
(default
No
)
R Limit
(default
No
)
Out of Tol
(default
No
)
ITE Limit
(default
No
)
TEC Open
(default
Yes
)
For each condition, if the menu item is set to
Yes
and the condition is detected,
the output will be shut off.
Temperature limits are a special case due to the multiple sensor inputs. There
are two menu items that control how the temperature limits are applied,
Apply
Nonactive
and
Apply Aux
. See
Applying Sensor Limits
below for more details
on these settings.