background image

6300 Series 

ComboSource

 User’s Manual · Page 57 

 

 

disabled output 

forced a shutdown of the laser output. 

E-512 

Power failure 

A power failure was detected. 

E-514 

Laser mode change 
disabled output 

A change in the operating mode of the 
ComboSource while the output was on shutdown 
the output. 

E-516 

Incorrect configuration 

for calibration to start 

The ComboSource was not configured properly, 

including the mode and output on state, to be 
able to start the desired calibration process. 

E-517 

Calibration must have 
the output on to start 

The laser output must be on for the calibration 
process to start. 

E-521 

TEC temperature limit 
disabled output 

The laser output was turned off because the TEC 
temperature limit was exceeded corresponding 

bit in the OUTOFF register was set. 

E-534 

Po mode selected with 
PD Response set to 
zero 

Attempted to select Po mode and PD Response 
was zero, or ComboSource was in Po mode and 
PD Response was set to zero. 

E-535 

Calibration cancelled 

The active calibration process was cancelled. 

E-536 

Intermittent contact 

fault 

The instrument detected an intermittent contact 

and shut down the laser output. If this is 
triggering falsely (such as in a noisy 
environment), the intermittent contact detection 
can be disabled in the main menu. 

E-537 

Thermal trip 

Excessive power dissipated inside unit. Lower 
voltage limit or add series resistance. See 

“Thermal Considerations” for more details. 

W-800 

Remote voltage sense 
is low 

When in remote voltage sense mode, the  
ComboSource detected a significant difference 
between the remote voltage and the voltage at 
the connector, which may indicate a problem with 
the remote voltage sense connection.   

W-801 

Burst Mode, Hold 
Output 

When in Io (Burst) mode, to turn the output on, 
the Output button must be held down for at least 
one second. If it is held down for less than one 
second, this warning message informs the user 
than the Output button press did not turn the 
output on. 

E-998 

Command not 
supported 

A command was recognized but not supported 
by the ComboSource. 

E-999 

Non-specific error 

A non-specific error was encountered. 

 

 

 

Summary of Contents for ComboSource 6300 Series

Page 1: ......

Page 2: ...Laser Driver 31 Laser Control Modes 31 Using the Laser Limits 32 Analog Modulation 33 Compensating for Voltage Loss 34 Using the Temperature Controller 37 TEC Control Modes 37 Using the TEC Limits 37...

Page 3: ...omboSource also operates in constant voltage control mode allowing you to use it in precision voltage applications such as EML devices or for doing V I measurement graphs The user interface of the Com...

Page 4: ...ldering into custom solutions is available as p n 1221B 1260B TECSource Cable 5A 2m This cable has DB 15 male female connectors for interfacing to the LaserMount or other connectorized fixtures and su...

Page 5: ...ut important information If at any time any of the following conditions exist or are suspected of existing discontinue use of the unit until it can be inspected by qualified service personnel Visible...

Page 6: ...manual for instructions on how to obtain service for this instrument WARNING To avoid electrical shock ensure a 3 prong power cord is used and is plugged into a earth grounded receptacle Failure to do...

Page 7: ...knob again to save the value Scroll down and make the same adjustments to the voltage limit Vf Lim Next you should adjust the temperature control settings appropriate for your device Scroll back to t...

Page 8: ...to be reduced If the temperature is slowly moving up and down try a higher Gain You may need to experiment with several gain settings to find the ideal value and for even finer control you can set the...

Page 9: ...ng a three to two prong adapter and connecting the ground tab to earth ground Turn the power switch located on the front panel into the on position The unit will display the model serial number and fi...

Page 10: ...oom ambient does not exceed the unit s operating temperature When using 1401 RM 2 to rack mount two units either unit may also be a 4300 Series LaserSource or 5300 Series TECSource Warm up and Environ...

Page 11: ...uts are controlled via the two On Off buttons For the laser when turning the output on there is a programmable delay factory default of 3 seconds where the unit will indicate that the output is on the...

Page 12: ...found in the Error Messages section below When the unit is in remote mode the yellow Remote LED will be lit More information about how the instrument behaves in remote mode can be found in the Remote...

Page 13: ...f the ComboSource is its ability to display both the set point and measurements simultaneously when in small display mode The table below shows which values will appear on the display depending on the...

Page 14: ...more information on the various modes Status Messages The instrument will display status messages in upper right corner indicating several different conditions for the laser that may be of interest t...

Page 15: ...he Control Modes section above the unit offers seven control modes Io ACC Io HiBW ACC Im AMC Po APC and Vf AVC Change this setting to select a new mode Io ACC Range Selects the high or low current ran...

Page 16: ...ble and connector resistances Cable R is ignored in Vf AVC Mode For more information about this feature see Using the Cable R Setting below Not available when using remote voltage sense 0 0000 Int Con...

Page 17: ...d User Cal Edit Cal Must be set to Yes before user calibration settings can be changed No Reset Cal Reset the Laser user calibration data to factory defaults all slopes to 1 all offsets to 0 Requires...

Page 18: ...in T Mode and the actual temperature drops below this value the output will be turned off Temperature limits not monitored in R Mode or Ite Mode Limit monitoring can also be disabled in software see T...

Page 19: ...information Off Ext Fan Pwr When Ext Fan is set to Custom this sets the specific fan voltage from 4 0 to 12 0 volts See the External Fan Control section below for more information 12 0 Ext Fan Mode C...

Page 20: ...Menu Advanced Invert ITE Reverses the direction of the current No Off if Thi Lim Turns off the TEC output if the TEC exceeds the temperature high limit Yes Off if Tlo Lim Turns off the TEC output if t...

Page 21: ...t While Rmt To disable changing the set point while in Remote Mode set this value to No To enable changing the set point while in Remote mode set this Value to Yes No Err While Rmt To turn off the dis...

Page 22: ...figurations Using the Main Menu System Menu Config menu item or the SAV and RCL commands over the computer interfaces ComboSource allows you to save up to four configurations Each configuration will s...

Page 23: ...ent Any command except DELAY or WAI commands can be used to construct a script allowing you to configure the instrument change modes change set points turn outputs on or off or any other of a wide ran...

Page 24: ...A C button and likewise the primary B and secondary D actions both use the same B D button To initiate the action assigned to the A or B keys press and hold the A or B key until the instrument does a...

Page 25: ...drive circuit and photodiode measurement circuit A cable is required for both the TEC and Laser outputs Arroyo Instruments carries cables specifically designed for these applications both with a DB c...

Page 26: ...ale DB 9 and has the following pin out DB 9 Pin Description 1 Interlock 2 Interlock 3 Earth Ground 4 Laser Cathode Voltage Sense 5 Laser Cathode 6 Photodiode PD Cathode 7 Photodiode PD Anode 8 Laser A...

Page 27: ...rom any modulation source The center pin is positive and the outer shell is ground TEC Output Connector The TEC Output connection is a female DB 15 and has the following pin out DB 15 Pin Description...

Page 28: ...and cathode connections are also isolated from earth ground However if you use the earth ground pin of the Output connector it is possible to create a ground loop if the instrument s earth ground is c...

Page 29: ...the ComboSource In the event you have multiple Arroyo Instruments products plugged in simultaneously you will need to experiment to see which instrument was assigned to which port For example you coul...

Page 30: ...dd a series resistor to the laser or TEC as appropriate to remove some of the power from inside the unit Typically a 30W 1 load resistor in series with the output is sufficient but higher resistances...

Page 31: ...control modes are used to drive a specific current through the laser diode When in this mode the set point will be in milliamps and the ComboSource will drive the desired current through the laser dio...

Page 32: ...ion BNC on the front panel of the instrument Modulation rates vary by model so see your model s specification for the maximum modulation rates Only Io HiBW mode supports high speed modulation All othe...

Page 33: ...function is 50mA V In addition the set point adds to the modulation input so using the same 6305 example if you were to apply 5V and the set point was 100mA then the resulting drive current would be 3...

Page 34: ...no special wiring to implement Four wire sense on the other hand does require additional wiring but is the most accurate method Details on both are found below Using Remote Voltage Sense The ComboSou...

Page 35: ...tting The Cable R setting allows you to calculate the voltage at the laser by subtracting the voltage loss through the cable and connectors This is done by measuring or calculating the cable resistanc...

Page 36: ...age is also the resistance of the cable To short the cable disconnect the laser and short the anode and cathode together as close to the end of the cable as possible Ideally the short should be done b...

Page 37: ...nted in software but due to the relatively slow nature of the TEC software is typically adequate Both high and low temperature limits are available and will turn the TEC output off unless this has bee...

Page 38: ...ture increases hence the negative coefficient Here is a typical response curve of a thermistor Typical Resistance vs Temperature Graph As can be seen be the graph the resistance of the thermistor drop...

Page 39: ...lly be obtained from the thermistor manufacturer The ComboSource defaults to the coefficients for the BetaTHERM 10K3A1 thermistor A 1 12924x10 3 B 2 34108x10 4 C 0 87755x10 7 You can change the coeffi...

Page 40: ...easurement mode which is subject to this error and a four wire measurement mode that uses separate sensor and source lines to remotely sense the actual resistance of the RTD and eliminate the cable or...

Page 41: ...temperature in C calculated using the standard conversion value of 1 C A or 0 1 C mV as appropriate and adjusted from Kelvin to Celsius External Fan Control The ComboSource has a built in user adjust...

Page 42: ...nal heating If your application requires you can also configure the ComboSource to operate in a cool only mode This is only possible with Peltier coolers as resistive heaters will heat regardless of t...

Page 43: ...the ComboSource will step through a process to determine the thermal response of the mount which can then be used in a mathematical model to calculate the PID parameters The PID parameters generated...

Page 44: ...tep it is performing and the word AutoTune will flash on the display The typical temperature profile of an AutoTune process is shown in the graph below The first step which is the calculation of the P...

Page 45: ...nchanged an E 436 will also be displayed Using the AutoTune Function Remotely The AutoTune function can be done remotely and works essentially the same as local operation The TEC AUTOTUNE command star...

Page 46: ...to short the terminals is not itself adding resistance to the circuit Use short heavy gauge wires preferably soldered onto the terminals to minimize any resistance the short might add Place the short...

Page 47: ...gh the User Cal menus or via the LASER USERCAL or TEC USERCAL commands over the computer interface For calibration of the instrument from the front panel the User Cal menu is located under the Advance...

Page 48: ...nt slope V Rem B Vf remote sense measurement offset V TEC User Calibration Entries Menu Entry Description ITE Set M ITE set point slope ITE Set B ITE set point offset A ITE M ITE measurement slope ITE...

Page 49: ...fset value for the high range Io set point which is the I Set HiB menu entry Considering that you need to raise the set point by 1 4mA then the I Set HiB menu entry should be adjusted to 1 4mA the I S...

Page 50: ...get a compensation that increases the set point by slightly over 1 4mA to bring it up to 300mA Let s check 300mA test point 1 0147 300 2 9934 301 4166mA check Likewise the 100mA set point should be l...

Page 51: ...ense resistor so both much be considered when determining the overall accuracy of your measurement system Considerations for Calibrating the TEC Sensors For calibration of resistance sensors thermisto...

Page 52: ...URRENT Range A 2 5 000 Resolution A 0 1 Accuracy set A 0 05 1 Stability ppm time 200 24 hours Temperature Coeff ppm C 200 PD Bias V 0 to 5V programmable LASER VOLTAGE Range V 0 10 0 10 0 10 0 4 Resolu...

Page 53: ...0 1 Accuracy set A 0 05 1 Stability ppm time 200 24 hours Temperature Coeff ppm C 200 PD Bias V 0 to 5V programmable LASER VOLTAGE Range V 0 18 0 15 Resolution V 0 001 Accuracy set V 0 05 0 005 Stabi...

Page 54: ...THERMISTOR Range k 0 2 450 Resolution k 0 01 Accuracy read k 0 05 50 100 A THERMISTOR Range k 0 02 45 Resolution k 0 001 Accuracy read k 0 05 5 LM335 Bias mA 1 Range mV 1730 4730 Resolution mV 0 1 Ac...

Page 55: ...ly E 201 Data out of range The message attempted to set a value that was outside the allowable range USB Serial only E 202 Invalid data type When trying to parse the message the data was in an invalid...

Page 56: ...determined manually E 501 Interlock shutdown output The interlock input pins 1 and 2 of the output connector were not shorted when the output was on E 504 Laser current limit disabled output The laser...

Page 57: ...535 Calibration cancelled The active calibration process was cancelled E 536 Intermittent contact fault The instrument detected an intermittent contact and shut down the laser output If this is trigge...

Page 58: ...upport Service Service and repair for the ComboSource can be obtained by contacting the distributor from where you purchased the instrument or directly from Arroyo Instruments A complete list of distr...

Page 59: ...e Test Standards EN 61010 Electrical Equipment for Measurement Control and Laboratory Use Safety Requirements This Certificate is the Manufacturer s Declaration which states that the 6300 Series Combo...

Page 60: ...Page 60 6300 Series ComboSource User s Manual Copyright 2019 Arroyo Instruments All Rights Reserved P N 530 1029F...

Reviews: