Page 48 · 6300 Series
ComboSource
User’s Manual
Laser User Calibration Entries
Menu Entry
Description
I Set LoM
Io set point, low range, slope
I Set LoB
Io set point, low range, offset (mA)
I Set HiM
Io set point, high range, slope
I Set HiB
Io set point, high range, offset (mA)
Im Set M
Im set point, slope
Im Set B
Im set point, offset (
μ
A)
V Set M
Vf local sense set point, slope
V Set B
Vf local sense set point, offset (V)
V Set Rm M
Vf remote sense set point, slope
V Set Rm B
Vf remote sense set point, offset (V)
I LoM
Io measurement low range, slope
I LoB
Io measurement low range, offset (mA)
I HiM
Io measurement high range, slope
I HiB
Io measurement high range, offset (mA)
Im M
Im measurement, slope
Im B
Im measurement, offset (
μ
A)
V M
Vf local sense measurement, slope
V B
Vf local sense measurement, offset (V)
V Rem M
Vf remote sense measurement, slope
V Rem B
Vf remote sense measurement, offset (V)
TEC User Calibration Entries
Menu Entry
Description
ITE Set M
ITE set point, slope
ITE Set B
ITE set point, offset (A)
ITE M
ITE measurement, slope
ITE B
ITE measurement, offset (A)
VTE M
VTE measurement, slope
VTE B
VTE measurement, offset (V)
Th 10
μ
A M
10
μ
A Thermistor measurement, slope
Th 10
μ
A B
10
μ
A Thermistor measurement, offset (k
Ω
)
Th 100
μ
A M
100
μ
A Thermistor measurement, slope
Th 100
μ
A B
100
μ
A Thermistor measurement, offset (k
Ω
)
LM335 M
LM335 measurement, slope
LM335 B
LM335 measurement, offset (mV)
AD950 M
AD590 measurement, slope
AD590 B
AD590 measurement, offset (
μ
A)
RTD M
RTD measurement, slope
RTD B
RTD measurement, offset (
Ω
)
RTD4 M
RTD 4-wire measurement, slope
RTD4 B
RTD 4-wire measurement, offset (
Ω
)
When the slope is 1 and the offset is zero, this is considered an
“uncompensated” measurement. If you change either the slope or offset, the
value becomes “compensated” based on these values. The compensation
formula (a linear equation) is very simple:
Compensated = Slope * Uncompe Offset