6300 Series
ComboSource
User’s Manual · Page 17
Menu
Description
Factory
Default
On Delay
The delay, from the time the
Output
button is
pressed to when the output is actually energized.
3000ms
Laser Menu
»
Advanced
Off if TEC
Off
Turns off the laser output if the TEC turns off
No
Off if TEC
Tlim
Turns off the laser output if the TEC exceeds the
temperature limit
No
Off if Io Lim
Turns off the laser output if the Io limit is active
No
Off if Im Lim
Turns off the laser output if the Im limit is exceeded
Yes
Off if Po Lim
Turns off the laser output if the Po limit is exceeded
Yes
Off if OutTol
Turns off the laser output if the laser goes out of
tolerance
No
Laser Menu
»
Advanced
»
User Cal
Edit Cal
Must be set to
Yes
before user calibration settings
can be changed.
No
Reset Cal
Reset the Laser user calibration data to factory
defaults (all slopes to 1, all offsets to 0). Requires
Edit Cal
to be set to
Yes
before this setting will be
displayed.
No
Various
M & B
Settings
Slope (M) and Offset (B) user calibration settings
for each set point and measurement. See the
User
Cal - Field Calibration
section below for additional
details on using slope/offset calibration.
No
TEC Menu
Mode
This sets the operating mode (
T Mode
,
R Mode
, or
Ite Mode
) of the temperature controller.
T Mode
Mount
Specify the mount connected to the
ComboSource
. If using a
LaserMount
, select the
appropriate model. Otherwise, select
User Defined
.
User
Defined
I Lim
I Lim
sets the current limit of the temperature
controller. The limit should be set to a value that is
suitable for your Peltier device.
3A
Summary of Contents for ComboSource 6300 Series
Page 1: ......