Description
TEC & General Specifications
TEC
SETPOINT
TEMPERATURE
Range (ºC)
1
-99 to 250
Resolution (ºC)
0.01
Therm Accuracy (± ºC)
2
0.05
3
AD560 Accuracy (± ºC)
2
0.05
LM335 Accuracy (± ºC)
2
0.05
RTD Accuracy (± ºC)
2
0.05
Stability (1hr) (± ºC)
4
0.004
Stability (24hr) (± ºC)
4
0.01
CURRENT
Range (A)
5
Compliance Voltage (V)
12
Max Power (W)
60
Resolution (A)
0.01
Accuracy (±[% set+mA])
0 + 30
Noise/Ripple (mA, rms)
< 5
MEASUREMENT
CURRENT
Resolution (mA)
10
Accuracy (±[% read+mA])
0 + 30
VOLTAGE
Resolution (mV)
10
Accuracy (±[% read V])
0 + 0.05
10
μ
A THERMISTOR
Range (k
Ω
)
0.2 – 450
Resolution (k
Ω
)
0.01
Accuracy (
±
[% read+k
Ω
])
0.05 + 50
100
μ
A THERMISTOR
Range (k
Ω
)
0.02 – 45
Resolution (k
Ω
)
0.001
Accuracy (
±
[% read+k
Ω
])
0.05 + 5
LM335
Bias (mA)
1
Range (mV)
1730 – 4730
Resolution (mV)
0.1
Accuracy (
±
[% read+mV])
0.3 + 1
AD590
Bias (V)
4.5
Range (
μ
A)
173 – 473
Resolution (
μ
A)
0.01
Accuracy (
±
[% read+
μ
A
])
0.03 + 0.1
RTD
Range (
Ω
)
20 – 192
Resolution (
Ω
)
0.01
1
Software limits. Actual range dependent on sensor type and system dynamics
2
Accuracy figures are the additional error the 5300 adds to the measurement, and does not include the
sensor uncertainties.
3
25°C, 100
μ
A thermistor
4
Stability measurements done at 25°C using a 10k
Ω
thermistor on the 100
μ
A setting. The number is ½
the peak-to-peak deviation from the average over the measurement period.