Page 10 · 6300 Series
ComboSource
User’s Manual
Rack Mounting
Rack mounting kits (
p/n 1401-RM-1
or
p/n 1401-RM-2
) for standard 19” racks
are available for the
ComboSource
, and supports the rack mount of one (
1401-
RM-1
) or two (
1401-RM-2
) units in a 2U (3.5”) high opening. Because the unit
draws air from the side, and therefore inside the rack housing, be sure that the
internal rack ambient temperature (which will typically be several degrees higher
than room ambient) does not exceed the unit’s operating temperature. When
using
1401-RM-2
to rack mount two units, either unit may also be a
4300 Series
LaserSource
or
5300 Series
TECSource
.
Warm-up and Environmental Considerations
In order to achieve the highest level of accuracy, the
ComboSource
should be
powered on for at least one hour prior to taking measurements. In addition,
ensure that the unit is not operating outside the operational temperature range
or humidity conditions.
CAUTION
Do not operate the unit above +40°C ambient, and
ensure the instrument is properly ventilated, or the unit
may overheat and possible damage to the unit may
occur.