No.
Message
Trigger condition
Info
Corrective
measure
-
No line
voltage.
Continue
with battery
supply?
no connection to mains
check mains
connection
connect device
to mains or
confirm
operation on
battery power
-
Battery
defect
Continue
without
battery
battery not ready for
operation
-
confirm
continued
operation
without battery
0
Sensor
defect Call
service
offset of the pressure
sensors outside
tolerance range
restart device
if error message
recurs, contact
authorised
service
technician
1
Sensor
defect Call
service
communication with
pressure sensors not
possible
restart device
if error message
recurs, contact
authorised
service
technician
2
Check
output! Keep
output open
while
switched on
pressure on tubing
system during switch-
on
disconnect
measuring line
from device and
restart device
no pressure on
tubing system
during switch-on
press 'Repeat'
3
EEPROM
defect Call
service
incorrect EEPROM
checksum
do not use the
device
contact
authorised
service
technician
4
O2 sensor
defect.
Continue
without
O2 sensor?
O
2
sensor not ready for
operation
-
contact
authorised
service
technician
5
Touch
screen not
calibrated
touch screen calibration
cannot be found in
EEPROM
calibrate touch
screen
contact
authorised
service
technician
Summary of Contents for AquaVENT FD140
Page 1: ...AquaVENT FD140 Instructions for use CRITICAL CARE...
Page 6: ...0 Contents 4 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 11: ...Instructions for Use AquaVENT FD140 Rev 12 9 2Safety In this chapter 2 1 Safety precautions 10...
Page 16: ...2 Safety 14 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 56: ...5 Operation 54 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 70: ...7 Cleaning 68 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 76: ...8 Maintenance 74 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Page 95: ...Instructions for Use AquaVENT FD140 Rev 12 93 This page is intentionally blank...