4 Preparing the device for operation
28
Instructions for Use, AquaVENT
®
FD140, Rev. 12
4.6 CPAP or Paed CPAP with face mask, heated
An active humidifier must be integrated in the heated CPAP breathing
circuit.
Note
To prevent water from the humidifier, which collects during
operation, from entering the AquaVENT
®
FD140, the active
humidifier should be placed at a level below that of the
AquaVENT
®
FD140.
Note
When removing the active humidifier, first disconnect the tube
from the AquaVENT
®
FD140 to prevent water from the
humidifier from entering the AquaVENT
®
FD140.