Setup Required
The Setup Required warning and alarm are
activated if the pump detects a fault in the
tubeset.
For both conditions:
• Switch off the pump.
• Call the service engineer.
If the pump is in Standby when the faulty
tubeset is detected:
• A warning is activated.
Screen A alternates with screen B, and
shows a faulty tubeset 1.
Therapy can not be started.
If the pump is in the Run state when the faulty
tubeset is detected:
• An alarm is activated.
Screen C alternates with screen D, and
shows a faulty tubeset 1. Therapy can
continue.
• When the operator next stops the pump
and puts it into Standby, the alarm reverts
back to the warning in which screen A
alternates with screen B, and shows a
faulty tubeset 1.
Therapy can not be restarted.
Service Required
The “spanner” icon appears at the lower left
side of the screen to indicate that the pump
requires service attention. This will normally
occur after the preset service interval has
expired.
Therapy can continue until service is
available.
Call the service engineer.
Condition
Description and Corrective Action
!
Setup required
Faulty
Tubeset
C
D
Setup required
!
Setup required
Setup required
A
B
Stop
130
Service “Spanner” icon