8
Where the tubesets enter the side of the pump,
the top tubeset is marked “1”.
2. Pressure indication: this is the inflation pressure
applied to the garment, as follows:
• Foot Garment: 130 mmHg.
• Calf and Calf & Thigh Garments: 40 mmHg.
3. Mains power indications:
4. Battery indication:
• When the pump is connected to the mains power
supply, then:
•• If the battery is fully charged, the indicator
will be static and show “full”.
•• If the battery is not fully charged, the indicator
will “scroll” from “empty” to “full” to indicate
that the pump is being charged.
• When the pump is NOT connected to the mains
power supply, then the indicator is static and
shows the charge remaining in the battery:
The battery has a service life of 5 years (600 charge
cycles). It is not user replaceable and must be
replaced as part of the service procedure.
5. Patient Hours Meter: if this is selected, the total
therapy time is shown (in hours). Refer to
“Changing Options” on page 16 to select and/or
clear the Patient Hours Meter.
• This is the pump run time since the Patient Hours
Meter was last cleared.
• The factory default is to NOT show the Patient
Hours Meter on the LCD screen.
6. Warning and alarm indications (refer to
“Troubleshooting” on page 22).
The pump is connected to the mains power supply.
The pump is not connected to the mains power supply,
but is powered by the internal battery only.
Battery Empty
Battery 1/3 Full
Battery 2/3 Full
Battery Full